General Instructions
Before First Use
• Remove all protective caps, sleeves, films, and packaging materials.
• Wash and inspect each instrument carefully before sterilization.
• Ensure hinged instruments are opened during cleaning to allow complete access to all surfaces.
• Verify that the instrument is free from manufacturing residues before use.
After Use
• Whenever possible, transport instruments dry to the reprocessing area.
• If wet transport is required, place instruments in an approved cleaning disinfectant solution.
• Completely dismantle instruments designed for disassembly before cleaning.
• Position instruments in cleaning trays or wire baskets so all surfaces are exposed.
• Rinse thoroughly with clean running water before further processing.
• Use ultrasonic cleaning only when appropriate and according to the equipment manufacturer’s instructions.
• Delicate microsurgical and fine-tipped instruments should be cleaned manually or placed in specially designed instrument holders.
Manual Cleaning & Disinfection
• Ensure all joints, hinges, lumens, channels, cavities, and openings come into contact with the solution.
• Follow the chemical manufacturer’s recommended concentration, temperature, and contact time.
• Use soft nylon or synthetic brushes to remove remaining debris.
• Never use steel wool, abrasive pads, or metal brushes.
• Clean internal channels using appropriately sized soft brushes.
• Rinse thoroughly using distilled or deionized water.
• Dry completely with a soft, absorbent, lint-free cloth.
• Dry internal channels and lumens using clean compressed air.
Machine Cleaning & Disinfection
• Follow the washer-disinfector manufacturer’s instructions.
• Position instruments so that cleaning solutions can reach every surface.
• Keep hinged instruments open during the cleaning cycle.
• Use only approved detergents and cleaning chemicals.
• Perform the final rinse using deionized or purified water.
• Allow sufficient time for complete drying.
• Remove instruments promptly after the cleaning cycle is complete.
Care, Testing & Maintenance
• Apply a medical-grade, steam-sterilizable instrument lubricant to joints, hinges, locks, and other moving parts.
• Inspect every instrument after cleaning and disinfection.
• Confirm that the instrument is clean and functioning correctly.
• Check for bent, cracked, fractured, corroded, loose, or worn components.
• Check cutting edges, jaw alignment, ratchets, and fine working tips.
• Remove damaged or defective instruments from service immediately.
Storage
• Place fine-tipped and microsurgical instruments in suitable protective racks or holders.
• Use dedicated instrument trays to prevent impact and surface damage.
• Secure ratcheted instruments only in the first locking position.
• Do not place heavy instruments on top of delicate instruments.
Sterilization
• Ensure instruments are completely clean and dry before sterilization.
• Do not overload sterilization trays or pouches.
• For pre-vacuum steam sterilization, a commonly used cycle is 134°C (273°F) with a minimum holding time of 5 minutes, subject to the instrument and sterilizer manufacturer’s instructions.
• Allow instruments to cool naturally before handling or storage.
• Always follow your healthcare facility’s validated sterilization protocol.
Important Safety Notice
• Never use damaged, corroded, or defective instruments.
• Use only approved cleaning agents, disinfectants, lubricants, and sterilization procedures.
• Follow applicable infection-control requirements and healthcare facility protocols.
• Proper care and maintenance help preserve instrument safety, precision, and service life.